+86 10-53676895
1368690224@qq.com
Routine maintenance of biochemical analyzers centers on cleaning, calibration, quality control, fluid system care, environmental stability, and record-keeping to ensure accurate, reliable results and prolong instrument life. Below are the key points organized by cycle and system:
1. Daily Maintenance
- Surface & Exterior Cleaning
Wipe the exterior, sample/reagent trays, and touch panel with 75% ethanol or mild detergent; remove spills, dust, and sample residues immediately.
- Probe & Needle Care
Clean the outer walls of sample/reagent probes and stirring rods with alcohol swabs; run automatic flushing to prevent clogging and cross-contamination.
- Fluid System Check
Empty and clean the waste container; confirm smooth drainage; verify sufficient purified water and cleaning solutions.
- Startup Checks
Confirm stable voltage, UPS function, temperature (37±0.1℃ for reaction bath), and humidity; run daily quality control (QC) and blank tests.
- Shutdown Procedure
Execute full internal cleaning cycles; cover reagents and store properly; log operation status, water consumption, and waste volume.
2. Weekly Maintenance
- Deep Cleaning of Reaction Cuvettes
Perform cuvette cleaning/soaking with alkaline or enzymatic detergent; ultrasonic cleaning to remove protein/fat deposits.
- Cuvette Blank & Integrity
Measure cuvette blank absorbance; replace scratched, cloudy, or damaged cuvettes.
- Tubing & Pump Maintenance
Inspect peristaltic pump tubes, seals, and connections for cracks/leaks; clean filters and nozzles.
- Mechanical Parts
Lubricate bearings and tracks lightly (medical-grade lubricant); remove dust from moving parts.
3. Monthly/Quarterly Maintenance
- Optical System Calibration
Calibrate wavelength, light source intensity, and absorbance; replace aging lamps; verify photometric accuracy.
- Temperature System Verification
Check incubator/reaction bath temperature with a calibrated thermometer; adjust if deviation >±0.3℃.
- Precision & Carry-Over Tests
Verify sample/reagent volume accuracy (error ≤±1% for small volumes); test cross-contamination levels.
- Filter & Sensor Checks
Clean/replace air filters, waste tank filters; check sensors for dirt or damage.

4. Key System-Specific Points
- Fluid & Wash System
Use only manufacturer-recommended cleaners; avoid generic agents that cause corrosion.
Regularly flush lines to prevent salt/protein buildup; replace worn tubing promptly.
- Sample/Reagent Handling
Keep reagents refrigerated after use; check expiration dates; avoid condensation dilution.
- Environmental Control
Maintain stable temperature (18–28℃), humidity (40–60%); avoid direct sunlight, dust, and strong electromagnetic fields.
- Software & Firmware
Keep OS, application, and firmware updated; back up data regularly.
- Safety & Documentation
Wear PPE; handle biohazardous waste properly.
Record all maintenance, calibration, QC results, repairs, and part replacements for audit and traceability.
5. Long-Term (Annual) Maintenance
- Full preventive service by qualified engineers.
- Replace worn components (lamps, pumps, seals, probes).
- Professional calibration and performance verification by metrology services.
Core Principle: Follow the manufacturer’s manual and strict schedules. Consistent preventive maintenance drastically reduces downtime, errors, and repair costs.
Contact below if you would like to get a reply quicker.