+86 10-53676895
1368690224@qq.com
Home All Products Biological Sample Preparation Technology > Nucleic Acid Extractors > High-throughput Nucleic Acid Extractor Made In China
The Auto-Pure4800 is a high-throughput fully automated nucleic acid extraction system. Adopting magnetic bead-based nucleic acid extraction technology, the instrument automatically completes all purification procedures within the workstation, including liquid transfer, lysis, washing, elution and PCR system setup. It enables one-stop operation with zero manual intervention. The maximum reaction volume is 1 mL, supporting simultaneous extraction of 48 samples each with a 1 mL reaction system. Automated workflows eliminate human-induced errors, delivering more reliable experimental data and superior assay reproducibility.
The Auto-Pure4800 is a high-throughput fully automated nucleic acid extraction system. Adopting magnetic bead-based nucleic acid extraction technology, the instrument automatically completes all purification procedures within the workstation, including liquid transfer, lysis, washing, elution and PCR system setup. It enables one-stop operation with zero manual intervention. The maximum reaction volume is 1 mL, supporting simultaneous extraction of 48 samples each with a 1 mL reaction system. Automated workflows eliminate human-induced errors, delivering more reliable experimental data and superior assay reproducibility.
Model | Auto-Pure 4800 Automatic Nucleic Acid Extractor |
Functions | Automatic sample loading, nucleic acid extraction and PCR reaction system setup |
Information Management | Supports external USB barcode scanners. The sample area is equipped with an integrated barcode scanning function, which automatically scans and imports data upon insertion of sample racks. |
Pipetting Principle | Air Displacement Pipetting Technology |
Pipette Quantity | Two single-channel pipettes (1000 μL), one single-channel pipette (200 μL) |
Pipetting Range | 1-1000μL |
Compatible Tips | Our brand tips, as well as domestic and imported tips compatible with this instrument |
Liquid Level Detection | The pipettor integrates pressure-based liquid level sensing, tip clogging detection and residual liquid volume detection functions |
Single Batch Sample Throughput | 1-48 samples per run |
Nucleic Acid Extraction Processing Volume | 50-1000μL |
Extraction Uniformity | CV≤5% |
Single-Sample Consumables | Supported |
Temperature Control for Extraction | Temperature control for elution and lysis: ambient temperature ~ 120°C |
Reagent Refrigeration Module | Optional accessory, temperature range: 0°C ~ 105°C |
UV Sterilization | Independent UV disinfection functions for extraction zone, pipetting zone and waste zone |
Fan Filtration System | Equipped with three HEPA high-efficiency filter screens |
Instrument Interface | USB port |
Power Supply | 220V,8A,50/60 Hz |
Power Consumption | Standby: 30 W, Maximum: 550 W |
Noise Level | ≤60dB |
Net Weight of Instrument | 130kg |
The bench support adopts a modular design, enabling expansion of various bench layouts.

The gas displacement pipetting principle prevents contamination of working reagents.Anti-droplet technology eliminates contamination caused by dripping.The negative pressure filtration system blocks aerosol contamination.Full-coverage UV disinfection achieves dead-angle-free cleaning of the workbench.

The graphical user interface allows program editing without complicated operations.A comprehensive built-in library of racks, consumables and liquid types cuts down the time for users to match programs.Operators can complete the full workflow of pipetting, extraction and system setup simply by ticking relevant options as required.

Equipped with a 1+2 pipettor combination, all pipetting channels adopt air-pressure liquid level sensing technology.Detailed aspiration parameters can be set for different liquids to significantly improve pipetting accuracy.