+86 10-53676895
1368690224@qq.com
Here are the key points for maintaining abiochemical analyzerto ensure accurate results, prolong equipment life, and minimize downtime:
1. Daily Maintenance
- Clean Sample & Reagent Probes: Wipe probes daily to prevent clogging and cross-contamination.
- Check Wash Stations: Ensure wash cups are clean and free of debris; confirm proper washing fluid levels.
- Inspect Sample/Reagent Carriers: Remove dust or spilled liquids from racks and trays.
- Run System Self-Check: Perform startup diagnostics to verify instrument status.
- Prime Fluid Systems: Flush water and cleaning solutions through tubing to maintain flow integrity.
2. Weekly Maintenance
- Deep Clean Probes and Pipettors: Use recommended cleaning solution to remove protein buildup.
- Clean Optics (if accessible): Gently wipe lens and light source (e.g., tungsten lamp) per manufacturer’s instructions.
- Check & Replace Filters: Inspect air and liquid filters; replace if dirty or clogged.
- Verify Liquid Levels: Refill cleaning solution, waste neutralizer, and rinse water as needed.
- Run Preventive Calibration Check: Use control serum to assess system accuracy.
3. Monthly Maintenance
- Replace Consumables: Change tubing, probe tips, or seals prone to wear.
- Calibrate the Analyzer: Perform full calibration using primary calibrators for critical tests.
- Inspect Pumps and Valves: Check peristaltic pumps and solenoid valves for leaks or irregular operation.
- Clean Internal Components: Access and clean sample/reagent discs, stirring rods, and reaction cups.
- Verify Temperature Control: Confirm incubator and reagent cooling unit temperatures are within spec.
4. Quarterly/Periodic Tasks
- Lamp Replacement: Replace light source (e.g., halogen/tungsten lamp) based on usage hours or intensity drop.
- System Performance Verification: Run comprehensive QC panels and compare with historical data.
- Software Updates: Install firmware or software upgrades as provided by the manufacturer.
- Electrical Safety Check: Inspect power cords, grounding, and internal connections.
5. General Best Practices
- Use High-Quality Water: Always use deionized or distilled water to prevent mineral deposits.
- Follow Manufacturer Guidelines: Adhere strictly to the user manual for procedures and intervals.
- Keep Detailed Logs: Record maintenance dates, part replacements, error codes, and calibration results.
- Train Operators Properly: Ensure all users understand correct operation and basic troubleshooting.
- Schedule Professional Service: Arrange annual or biannual preventive maintenance by certified engineers.
Regular and systematic maintenance ensures consistent analytical performance, reduces errors, and supports compliance with quality standards such as CLIA, ISO 15189, or CAP accreditation.
Contact below if you would like to get a reply quicker.