1. Installation & Placement
- Place the unit on a flat and stable bench. Reserve sufficient clearance for heat dissipation. Keep away from direct sunlight, heat sources and air‑conditioning outlets to avoid sharp ambient temperature fluctuations.
- Ambient operating temperature: 18‑26 ℃; ambient humidity ≤85%RH. Avoid dusty, humid or corrosive‑gas environments.
- Ensure reliable equipment earthing. Use an independent power socket; do not share power outlets with high‑power devices to prevent voltage fluctuation.
2. CO₂ Gas Supply
- Use high‑purity carbon dioxide (99.99%). Industrial‑grade CO₂ is prohibited to avoid cell contamination.
- Secure the gas cylinder against tipping. Fit a pressure‑reducing valve. Output pressure shall be set at 0.08‑0.12 MPa; excessive pressure is not allowed.
- Use dedicated aging‑resistant gas tubing. Perform regular gas‑leak checks. A gas filter is recommended to block impurities entering the chamber.
- Replace cylinders when gas pressure is low. Close the incubator inlet valve before replacement to prevent gas backflow.
- Minimize door opening/closing frequency to avoid drastic CO₂ concentration variation which compromises cell culture.
3. Temperature Control
- Standard setting: 37 ℃. Load samples only after temperature stabilizes; do not place specimens during heat‑up.
- Avoid over‑temperature operation. Do not over‑crowd the chamber; leave gaps between samples for uniform air circulation.
- Prevent liquid or heat sources from contacting internal sensors, which will cause temperature reading deviation.
4. Humidity Management (Humidity Water Pan)
- Fill the water pan with sterile distilled / deionized water only. Tap‑water is forbidden to prevent limescale and microbial growth.
- Check water level regularly; never run the pan dry. Replace water and clean‑disinfect the water pan periodically.
- High‑humidity conditions favour mould growth. Avoid spilled culture medium inside the chamber; keep sample vessels properly capped.
5. Operational Practices
- Open and close the chamber door quickly to limit loss of temperature, CO₂ and humidity; keep door‑opening time as short as possible.
- Clearly label all cultured samples. Wipe up spilled medium and liquid immediately to prevent contamination.
- Do not place unsealed liquid containers, which may cause excessive evaporation and spillage into air ducts.
- Install inner shelves securely. Do not apply heavy load to sensors or infrared detectors.
6. Cleaning, Disinfection & Maintenance
- Periodic chamber cleaning: power off and cool down the unit before wiping with 75% ethanol. Chlorine‑containing disinfectants such as 84 disinfectant are prohibited, as they corrode metallic components.
- For models with high‑temperature dry‑heat sterilization: remove water pan, sensors and plastic accessories before sterilization. Follow manufacturer procedures. Allow full cooling before resuming use.
- Replace HEPA filters according to specified service intervals to maintain internal air filtration performance.
- Perform regular calibration for temperature and CO₂ concentration. Keep infrared probes clean; dust will lead to inaccurate readings.
7. Safety & Troubleshooting
- Investigate alarms (over‑temperature, low CO₂ pressure, temperature fault) promptly. Do not continue incubating valuable samples under alarm conditions.
- For long‑term shutdown: drain the water pan, fully clean and dry the chamber, turn off CO₂ supply and cut off power.
- High internal humidity carries microbial‑hazard risk. Observe biosafety precautions during operation. In case of heavy mould contamination, complete thorough disinfection before reuse.
8. Prohibited Operations
❌ Tap‑water for humidity water pan
❌ Chlorine‑containing disinfectants for chamber wiping
❌ Unsecured gas cylinders or excessively high gas output pressure
❌ Overloaded chamber blocking air circulation
❌ Prolonged door‑opening
❌ Running sterilization cycle without removing sensors and plastic parts